BETSA
services
membrane diamond anvil cell
downloads
company
contact
photo betsa
BETSA
membrane diamond anvil cell
membrane diamond anvil cell

Pressure Ruby Luminescence [PRL]

Digital Camera USB.

Software V2.10.

---.

Laser.

Laser shutter.

Ruby.

PRL Accessories

---.

V-Bloc.

XYZ stage

Pressure Ruby luminescence.

Products

Pressure Ruby Luminescence.

All for very high pressure experiments

            The [PRL] spectrometer (Pressure by Ruby Luminescence) has been designed for measuring pressures using the luminescence of a reference material (Ruby or SrB407 : Sm 2+).

            Its small size and weight make "on site" measurements possible.When equipped with an internal mini laser it functions as a "stand alone" instrument.

Measurements are very easy and fast with custom high pressure calculation software.

Pressure Ruby Luminescence

Software V2.10

for high pressure measurement

WIN XP PRO



 

  

Digital Camera USB


High Sensitivity - Low Power Consumption - Compact Design




Features


CMOS Camera 1 / 3 " USB 2.0


Color or Monochrome mode

Dynamic 10-bit

752 x 480 pixels (WVGA)

Global Shutter

50 frames per second

High Dynamic Mode

high sensitivity


compact design


effective against the light and overexposure


Mounting: type C


Size | Weight: 50 mm x 29 mm diamtre (excluding C mount)


System Requirements: PC with USB 2.0; Windows 2000 SP4 Update Rollup 1; Windows XP SP2; Windows Vista SP1

Laser green or blue




Green laser module features


Wavelength: 532 nm


Output Power: 10 mW, CW, stabilized


Polarization: linear


Stability at constant housing temperature: <2% per 8 hrs


Beam diameter at output: 1 ± 0.2 mm


Beam divergence (half angle 1/e2): 0.6 ± 0.1 mrad


Mounting: on heat sink,




Blue laser module features

A solid state 473nm blue laser is made features of ultra compact, long lifetime, low cost and easy operating, which is used in experiment, fluorescence sensors.


Wavelength: 473±1 nm


Output Power: 500 mW,


Stability at constant housing temperature: <3% over 4 hrs


Beam diameter at output: 1 ± 0.2 mm


Beam divergence (full angle): <1. 5 mrad


Mounting: on heat sink,


TTL modulation/ Analog modulation




  

Laser shutter and white light

beam commutator [OBT01]


            This laser shutter is aimed at being installed on a BETSA PRL optical transfer.

It is designed to switch between the laser beam and a white light source, which allows for

            - Automatically block the laser beam when measuring the background signal (For safety purposes, the laser beam is “ON” only during the measurement step.)

            - Back lighting the sample when observing it.




Features


Entrance diameter for the laser beam : 5mm.


Entrance diameter for the white light : 7 mm.


Electronic remote control ( software or manual) :

0 V : laser off / light on.

5 V : laser on / light off.


Current : 20 mA.


Connection : Connected to PRL Interface .


Small size : 50 x 52 x 16 mm.


Weight : 125 gr.



opt01.pdf





  

RUBY BALLS Al2O3


Our ruby balls are made for applications in spectroscopy.



            Tested and Approved, our ruby balls are ideal and provide a comfortable use when measuring pressure by fluorecence. Pressure measurement performed with the spectrometer [PRL].



Physical properties:

- Mean diameter: 3 microns to 50 microns.

- Specific surface: 0.08 m² / g.

- Doping: 3600 ppm Cr3 +.

- Density: Apparent 2.3; packed 2.4 .


  

Pneumatic Drive System

for the MDAC high pressure cell

[PDS 151]


            The Pneumatic Drive System is used to control the force applied to the MDAC anvil by varying the pressure in its membrane.

            The small size, weight, and portability of the PDS allow its use in a variety of environments.



Features:


0 to 150 bars Pressure range.


Digital Pressure display with a 0.1 bar resolution.


Fine and continuous Pressure control with 2 valves ( " IN " and " OUT " )


High pressure connection for the output (on the front panel): Nova Swiss female for 1/16’’ capillary.


High pressure connection for loading (on the rear panel): Tube ¼’’ Gyrolock female.


Portable Size : 260 x 110 x 270 mm; Weight : 2,9 Kg.



pds151.pdf

[V-BLOC]



            V-Block is a piece to maintain the M DAC on the BETSA [XYZ3] stage .

It also possible to maintain various types of BETSA optical transfer.

 

            The standard size of the [V-BLOCK] is made to place Ø 50 mm parts.





It is possible to manufacture according to your needs others size of V-BLOCK.

  

3 axis (XYZ) [XYZ3].


            The [XYZ3] is a holder featuring a 3 axis stage. It is used for positioning the MDAC high pressure cells centering and focusing on experiment setups.




Features:


External sizes :

                H= 146 mm.

                L= 65 mm.

                Depth= 50 mm.


Travel :                                  Sensitivity :

                X= 6 mm.                               250μm / turn.

                Y= 6 mm.                               250μm / turn.

                Z= 8 mm.                               250μm / turn.

                

Load capacity :

800g.


High rigidity.



xyz3.pdf


  

Spectrometer for high pressure measurements

[PRL]


                The PRL spectrometer (Pressure by Ruby Luminescence) has been designed for measuring pressures using the luminescence of a reference material (Ruby or SrB407 : Sm 2+).

                Its small size and weight make "on site" measurements possible.When equipped with an internal mini laser it functions as a "stand alone" instrument.

Measurements are very easy and fast with custom high pressure calculation software.



Features


Resolution : 0.4 nm.


Spectral range : 675-745 nm.


Sensitivity : 0.005 nm.


Spectrometer /Microscope connection by fiber optics.


Optical transfer (microscope) : Magnification 400.


Dedicated software : for high pressure measurement (English / French versions).


Calibration : Very fast with an internal Neon source.


Portable equipment : Weight : about 11kg. (depending on options) ; External sizes : 33 x 10 x 30cm.



OPTIONAL ITEMS


Laser beam / light source beam switching box To be inserted between the laser and the optical

transfer.

High resolution small size video camera To replace the eyepiece.

XYZ stage To control the motion either of the MDAC or of the optical transfer.

Mini laser module (10 mW @ 532 nm) .



prl.pdf